We describe a reversed-phase HPLC method for determination of total mycophenolic acid (MPA), its free concentration (MPAf), and the glucuronide metabolite (MPAG), based on simple sample preparation and gradient elution chromatography. The compounds were quantified in parallel by absorbance at 254 nm and 215 nm in the internal standard mode. Linearity was verified up to 50 mg/L for MPA and up to 500 mg/L for MPAG (r >0.999). Detection limits at 215 and 254 nm were, respectively, 0.01 and 0.03 mg/L for MPA, and 0.03 and 0.1 mg/L for MPAG. The recovery of MPA was 95–106%;recovery of MPAG was 96–106%. The imprecision (CV) for MPA (0.2–25 mg/L) was <8.4% (254 nm) and <4.4% (215 nm) within day (n = 12) and <9.2% (254 nm) and <6.2% (215 nm) between days (n = 12). The imprecision for MPAG (10–250 mg/L) was <4.9% (254 nm) and <3.4% (215 nm) within day, and <6.1% (254 nm) and <5.9% (215 nm) between days. For quantification of MPAf, 100 μL of ultrafiltrate was applied directly to the column. The detection limit was 0.005 mg/L at 215 nm and 0.015 mg/L at 254 nm. In the range between 18–210 μg/L, the within-day CVs were <11.8% (n = 12) and the between-day CVs were <15.8% (n = 12).