Reflotron cholesterol measurement evaluated as a screening technique.

Author:

Sedor F A1,Holleman C M1,Heyden S1,Schneider K A1

Affiliation:

1. Hospital Laboratories, Duke University Medical Center, Durham, NC 27710

Abstract

Abstract We evaluated the analytical performance of Boehringer Mannheim Diagnostics' "Reflotron" analyzer for the measurement of cholesterol. Coefficients of variation (CVs) for whole-blood cholesterol were: within-day 2.0% and 2.2% at 1680 and 2670 mg/L, respectively; between-day 1.8% and 2.4% (n = 9 and 8). Results were similar for serum and heparinized or EDTA-treated single-donor plasma (CV 1.4% to 2.6%). CVs of results for two reconstituted commercial quality-control materials were 3.4% and 4.6%. Heparin and hematocrit were evaluated as interferents, and critical limits for interference were identified for bilirubin, hemoglobin, and triglyceride in blood and plasma or serum. When sample collection and analysis were controlled by trained personnel, results with the Reflotron (y) compared well with those by the Ektachem procedure (x) for both blood and serum samples: r = 0.950, y = 0.944x + 130 mg/L; and r = 0.955, y = 0.93x + 43.5 mg/L, respectively. The same comparability was observed when the analysis was performed by briefly trained high-school students: r = 0.980, y = 0.949x + 23 mg/L. Performance decreased when both collection and analysis were performed by laymen: r = 0.880, y = 0.870x + 186 mg/L.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

Cited by 9 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Determination of cholesterol in blood. Part 2;Review Journal of Chemistry;2016-04

2. Comparison of Total Cholesterol Measurement Using Reflotron Capillary Analysis and Laboratory Venous Blood Analysis;American Journal of Health Promotion;1995-09

3. The use of desk-top cholesterol analysers in general practice;Public Health;1995-09

4. Reflotron;Dry Chemistry - Analysis with Carrier-Bound Reagents;1993

5. Performance of the Reflotron and the Seralyzer in primary health care;Scandinavian Journal of Clinical and Laboratory Investigation;1993-01