Abstract
Abstract
Automated HPLC is used to test dried blood-spot specimens from newborns for hemoglobins (Hb) F, A, S, C, E, and D. We present the method and report on its performance determined during >4 years of testing 2.5 x 10(6) newborns. The method features automated derivation of presumptive phenotypes; quantitative quality control and proficiency testing; throughput of one specimen per minute; small sample volume; hemoglobin concentrations quantified with an interlaboratory CV of 14-18%; retention times with interlaboratory CV of <2% and matching, within +/- 0.03 min, of laboratories and reagent lots; control of peak resolution; 0.5% detection limit for Hb S and C, and 1.0% for Hb F, A, E, and D; few interferences; and negligible background and carryover. Shortcomings of the method are the absence of microplate barcode identification and the need for manually pipetting the sample eluate into the microplate.