Abstract
Abstract
I describe a new assay that is capable of measuring vitamin D2, vitamin D3, 25-hydroxyvitamin D2, and 25-hydroxyvitamin D3 in 2 ml of plasma or serum. Plasma is extracted by the Bligh and Dyer technique [Can. J. Biochem. Physiol. 37, 911 (1959)], the lipid component is fractionated by two high-performance liquid-chromatographic systems based upon adsorption and reversed-phase chromatography, and each of the four vitamin D metabolites is measured by its absorbance at 254 nm. The method has a sensitivity limit of 0.5 mug/liter of plasma. The identity of metabolite peaks was confirmed by mass spectrometry, ultraviolet absorption spectrophotometry, and rechromatography, and there was good correlation (r=0.84) between plasma 25-hydroxyvitamin D as measured by the present method and by a protein binding assay developed in our laboratory. Mean concentrations of vitamin D and 25-hydroxyvitamin D in normal adults (n=25) in December were 2.2 +/- 1.1 (SD) and 16 +/- 3.9 (SD) mug/liter, respectively. 25-Hyroxyvitamin D2 made up 31% of the total 25-hydroxyvitamin D. Patients receiving pharmacological doses of vitamin D had values for vitamin D and 25-hydroxyvitamin D that were 10- to 100-fold normal. This method provides a rapid, reliable physico-chemical assay that appears to have advantages over existing protein binding assays and can be used to measure circulating vitamin D.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
155 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献