Rapid, Accurate Technique for Determination of Carbon Monoxide in Blood

Author:

Dahms Thomas E1,Horvath Steven M1

Affiliation:

1. Institute of Environmental Stress, University of California, Santa Barbara, Calif. 93106

Abstract

Abstract We report a chromatographic technique in which a vortex method of gas extraction and a thermal conductivity detector are used to measure carbon monoxide in blood. This method has the advantages of a small sample volume (100 µl), a rapid analysis time (3 min), ease of operation, and a high correlation with results of the standard gasometric method over the entire possible range (r = 0.999). The accuracy of this method (coefficient of variation) for measuring CO in human blood samples was 1.7%. With it, dissolved CO can be accurately measured, as demonstrated by the Bunsen solubility coefficient for plasma CO by this method of 22.0 ml of CO per liter of water per atm. (101 kN/m2) at 24 °C (theoretical value: 21.8).