Evaluation of a New Procedure for Measuring Serum Creatine Kinase Activity

Author:

Henry Wilkinson J1,Steciw B1

Affiliation:

1. Pepper Laboratory of Clinical Medicine, Hospital of the University of Pennsylvania, Philadelphia, Pa. 19104, and Smith Kline Instruments, Inc., Palo Alto, Calif. 94306

Abstract

Abstract A new spectrophotometric microtechnique for the determination of serum creatine kinase activity, in which all reagents are provided in a single com-pressed tablet, has been evaluated. The procedure depends upon coupling the creatine phosphate-ADP reaction with the hexokinase and glucose-6-phosphate dehydrogenase reactions. The new technique is quick, relatively simple, and gives results which compare favorably with the conventional spectrophotometric method in precision and sensitivity. It requires a sample volume of 10 Al, and values ranging from 10 to1600 U/liter can be determined without dilution. Gross hemolysis leads to erroneously high values, but the error is negligible with slightly hemolyzed specimens. A provisional normal range has been established