Affiliation:
1. Pathology Department, The Swedish Hospital Medical Center, Seattle, Wash 98104
Abstract
Abstract
A new, atomic absorption spectrophotometric method for serum iron and iron-binding capacity is described, which utilizes 20% (w/v) TCA plus heating at 90° for 15 min. This procedure liberates the ferric iron; precipitates the protein, facilitating removal by centrifugation; and avoids significant interferences by contaminating hemoglobin iron. The method is highly specific, accurate, and has the additional feature of requiring less time than most colorimetric or atomic absorption methods employing chelation and extraction.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
81 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献