Background: The concentration of thiamin diphosphate (TDP) in erythrocytes is a useful index of thiamin status. We describe an HPLC method for TDP and its results in patients at risk of thiamin deficiency.
Methods: We used reversed-phase HPLC with postcolumn derivatization with alkaline potassium ferricyanide and fluorescence detection. Samples were deproteinized and injected directly onto a C18 column. TDP concentrations in erythrocytes were compared with those in whole blood. Reference intervals for erythrocyte TDP (n = 147; 79 males and 68 females; mean age, 54 years) and whole blood TDP (n = 124; 68 males and 56 females; mean age, 54 years) were determined in an apparently healthy population. We compared erythrocyte TDP with results of the erythrocyte transketolase activation test in 63 patients who were considered at risk of thiamin deficiency.
Results: The method was linear to at least 200 μg/L. The between-run CV was <8%. The lower limit of quantification for both whole blood and packed erythrocytes was 300 pg on column with a detection limit of 130 pg on column. Recovery of TDP from blood samples was >90%. TDP in erythrocytes correlated strongly with that in whole blood (r = 0.97). Reference intervals for erythrocyte and whole blood TDP were 280–590 ng/g hemoglobin and 275–675 ng/g hemoglobin, respectively. Of the 63 patients suspected of thiamin deficiency, 46 were normal by both TDP and activation tests, 13 were deficient by both tests, 1 was deficient by the activation test but had normal erythrocyte TDP concentrations, and 4 were normal by the activation test but had low TDP.
Conclusions: The HPLC method is precise and yields results similar to the erythrocyte activation assay.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
158 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献