Author:
Burkhardt A E,Johnston K G,Waszak C E,Jackson C E,Shafer S R
Abstract
Abstract
A solid-phase reagent for determination of urinary specific gravity (relative density) is described. This reagent strip, similar to others in the "N-Multistix" series (Ames), contains a polyacid whose acidity is sensitive to the ionic concentration in the urine in which it is immersed. As the acidity of the polyacid changes, pH changes are detected by a pH indicator within the reagent strip. In comparison studies, 84.4% of relative densities as measured with these reagent strips were within 0.005 of the corresponding results with a total-solids meter, and 89.9% were within 0.005 of the corresponding urinometer results. Adding a correction of +0.005 to the reagent-strip results for urines with high pH increased the percentage of results within 0.005 of the comparison method to 90.7% (TS meter) and 92.9% (urinometer). Lot-to-lot variability and reader-to-reader variability were both low. Reagent strip results are not affected by glucose, may be increased by albumin, and correlate with urea concentrations.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry