Bretaudiere J P,Phung H T,Bailly M
Abstract
Abstract
A direct enzymatic micromethod (sample volume, 3mul) has been adapted to the centrifugal analyzer (ENI-GEMSAEC) for measurement of urea in plasma and urine. The method is based on urease (urea amidohydrolase, EC3.5.1.5)/glutamate dehydrogenase [l-glutamate:NAD(P)+oxidoreductase (deaminating), EC1.41.3] coupled reactions, and uses a two-point fixed-time (t(1)=20s,t(2)=50s)kinetic scheme for monitoring the rate of comsumption of NADH at 340 nm. Sensitivity and precision of the method are excellent,and results compare well with those from a commonly used continuous-flow method.