Author:
Seccombe D W,Dodek P,Frohlich J,Hahn P,Skala J P,Campbell D J
Abstract
Abstract
Because of renewed interest in a possible connection between carnitine, lipid disorders, and myopathy, an automated method of analysis is desirable. Deproteinization of serum by use of membrane filter cones and automated assay with a bichromatic analyzer (the ABA-100) substantially increases efficiency without sacrificing the specificity and accuracy of the original manual enzymatic method. The described procedure allows for analysis of 80 speciments a day and is thus suitable for screening of selected populations. Normal values found in blood sera of adults were in the range of 25.0-73.8 mu mol/liter and the method has sufficient sensitivity to accurately measure concentrations as small as 10 mu mol/liter.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
25 articles.