Abstract
Abstract
Thymol blue monophosphate is evaluated as a substrate for use in kinetic bichromatic analysis of serum alkaline phosphatase activity. The effect of abnormally high concentrations of hemoglobin, bilirubin, or chylomicrons is negligible. The expected range for ambulatory adults established by assaying sera from 206 blood donors, was 12-44 U/lliter at 37 degrees C. Comparison of the method with a commonly used p-nitrophenyl phosphate method yielded a correlation coefficient, r, of 0.9949. Advantages of the method are the reduced cost per test and the exceptionally stable substrate.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
4 articles.