Affiliation:
1. Centre de Recerca Biomèdica, Hospital de Sant Joan de Reus, Spain
Abstract
Abstract
A considerable demand for convenient, rapid, inexpensive assays of ferritin in serum has been generated in recent years in hospital laboratories and blood banks. We describe a simple and rapid particle-enhanced turbidimetric immunoassay suitable for routine application in a Monarch 2000 centrifugal analyzer with commercially available reagents. This fully automated assay (y) requires no pretreatment of sample, and correlation with a two-step sandwich ELISA (x) is excellent (y = 1.018x + 0.397, Sy/x = 0.027). The analytical range extends from 5 to 900 micrograms/L. Intraassay imprecision (CV) ranged from 1.1% to 5% for various specimen concentrations. Interassay imprecision ranged from 2.2% for above-normal concentrations (755 micrograms/L) to 9.5% for low concentrations (39 micrograms/L). No specimen-related carryover was detected. The method has been useful in our predeposit autologous blood transfusion program for rapid assessment of iron status in patients undergoing repeated phlebotomies.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry