Author:
Vlahos I,MacMahon W,Sgoutas D,Bowers W,Thompson J,Trawick W
Abstract
Abstract
In this method, we use the Amicon MPS-1 centrifugal ultrafiltration device and the YMB membrane in measuring free testosterone in serum. Two independent assays are combined: total testosterone and the ultrafiltrable fraction of added [3H]testosterone. The unbound fraction is determined in 0.15-0.5 mL ultrafiltrates of 0.6 to 1 mL of variably diluted serum that has been equilibrated with [3H]testosterone at 37 degrees C. The assay is rapid (less than 1 h), practicable (requires 0.6 mL of serum), and reproducible (CV 3.2% within assay, 3.9% between assays). Accuracy was evaluated as the fraction of free testosterone in the ultrafiltrate of dialyzed serum vs that in a prior dialysate; they were the same confirming the validity of the free testosterone measurement. Samples from ostensibly healthy men and women and from hirsute and pregnant women gave results that agreed with those obtained by equilibrium dialysis. Total testosterone concentrations for normal and hirsute women showed considerable overlap, but data on free testosterone concentrations in these populations were better resolved.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
70 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献