Author:
Wessels R A,Rogers B B,Ou C N,Alcorn R,Buffone G J
Abstract
Abstract
"High-performance" liquid chromatography (HPLC), applied to hemoglobin analysis, is decidely more sensitive and gives better resolution than do routine electrophoretic methods. Here we present a case with a rare double heterozygote hemoglobin S/LeporeBoston, originally diagnosed as homozygous hemoglobin S by routine electrophoretic methods. Using a gradient elution weak cation-exchange HPLC technique, we could separate hemoglobin S and hemoglobin LeporeBoston and make the correct diagnosis. This case demonstrates how HPLC can be a useful adjunct to routine electrophoresis.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献