Author:
Richardson A P,Kim J B,Barnard G J,Collins W P,McCapra F
Abstract
Abstract
This simple solid-phase chemiluminescence immunoassay for measurement of progesterone in extracts of venous plasma has sensitivity and precision similar to that of conventional radioimmunoassay with use of a tritiated antigen. The labeled antigen, 11 alpha-progesteryl-2-succinoyltyramine-4-(10-methyl)-acridini um-9-carboxylate, and a monoclonal antibody to progesterone-11 alpha-succinyl-bovine serum albumin are incubated with a 100-microL aliquot of plasma extract (equivalent to 20 microL of plasma) and 50 microL of a suspension of an IgG fraction of a donkey antiserum to mouse immunoglobulins, covalently attached to cellulose particles. After the antibody-binding reaction (60 min at 4 degrees C), 1 mL of phosphate buffer is added to each tube, the tubes are centrifuged (5 min, 1500 X g), and the supernatant fluid is aspirated. The washing step is repeated and diluted hydrochloric acid (50 mmol/L, 50 microL) is added to the pellet. Luminescence is initiated by oxidation with dilute sodium hydroxide/hydrogen peroxide. The signal is integrated over 10 s. The light yield is inversely proportional to the progesterone concentration in the standard or sample.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
23 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献