Author:
Kabra P M,Wall J H,Blanckaert N
Abstract
Abstract
In this simple, precise, accurate, and specific isocratic liquid chromatographic procedure for determining cyclosporine, the cyclosporine is extracted from 1 mL of whole blood or from plasma, with 500 micrograms of cyclosporin D added per liter as internal standard, by elution from a Bond-ElutTM C18 extraction column with 300 microL of a mixture of ethanol and tetrahydrofuran. A 100-microL aliquot of the eluate, injected onto a cyano-phase analytical column, is eluted with a mixture of acetonitrile and pH 7.0 phosphate buffer at a flow rate of 1.0 mL/min and at 50 degrees C. Detection is at 210 nm. The chromatography is complete in less than 14.0 min. The method can measure less than 10.0 micrograms/L. Analytical recovery of cyclosporine added to whole blood ranged from 99 to 109% for concentrations up to 2000 micrograms/L. Between-run CVs ranged from 6.4 to 6.6%. None of numerous drugs and steroids tested interfered. Results by radioimmunoassay exceeded by 20 to 350% those measured by the present method.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
35 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献