Eight compact analysis systems evaluated for measuring total cholesterol

Author:

Gregory L C1,Duh S H1,Christenson R H1

Affiliation:

1. Department of Pathology, University of Maryland School of Medicine, Baltimore 21201

Abstract

Abstract We examined the analytical performance of eight compact systems for measuring total cholesterol: AccuMeter, Cobas Ready, Discovery f2, DT60, L-D-X, Reflotron, QCA, and Vision. We determined average bias at two decision levels, the mean absolute bias, and the percentage of results differing from the comparison method results by > 8.9% allowable total error limit for multiple reagent lots. Average bias was < 3% for all lots tested for AccuMeter, Discovery f2, and DT60, but > 3% for one or more lots or sample types tested with the other systems. Of results from each reagent lot, > 95% were within the 8.9% total error specifications with Discovery f2, DT60, and QCA, whereas the performance of L-D-X, Vision, and Reflotron depended on reagent lot and (or) sample type. Of all results from each lot tested with AccuMeter and Cobas Ready, > 5% exceeded the total allowable error limit. We determined imprecision for five systems: Cobas Ready, Discovery f2, and QCA had CVs < 3%, whereas CVs for AccuMeter and L-D-X were > 3% but < 5%.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry (medical),Clinical Biochemistry

Cited by 11 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3