Affiliation:
1. Department of Clinical Chemistry, National Cardiovascular Center, Osaka, Japan
Abstract
Abstract
We developed a kinetic assay for calcium in serum and urine, based on the activation of porcine pancreatic alpha-amylase (EC 3.2.1.1) with 2-chloro-4-nitrophenyl-alpha-maltotrioside as substrate. The kinetic generation of 2-chloro-4-nitrophenol, monitored at 405 nm, was proportional to the concentration of calcium in serum and urine. The assay was developed and evaluated with the Cobas Bio centrifugal analyzer. The average within-run and between-day imprecision (CV) was 0.96/1.26% and 1.07/1.63%, respectively, for serum calcium and 1.50/2.54% and 1.70/2.64%, respectively, for urine calcium. Results of the proposed method (y) correlated well with those determined by atomic absorption spectrophotometry (x): y(serum) = 1.005x + 0.028 mmol/L (Sy/x = 0.058, r = 0.974, n = 50), and y(urine) = 1.017x-0.115 mmol/L Sy/x = 0.30, r = 0.981, n = 25). We also present data showing that the method is highly sensitive, rapid, relatively free of interference, and amenable to automation.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献