Affiliation:
1. Serono Diagnostics, Ltd., Woking, Surrey, UK
Abstract
Abstract
Methodologies for establishing standard curves on automated immunoassay systems are dependent on the computational capability of the instrument, its throughput, and, perhaps, the technical experience of the user. Here, factors constraining this approach are discussed in general and the procedures used for the Serono SR1 analyzer are considered in detail. The SR1 is a moderate-throughput, fully automated immunoassay system capable of performing tests for (currently) 18 analytes. The associated user-generated standard curves are utilized not only for interpolation of results but also for evaluation of system performance. The choices of curve-fitting method, weighting of data, acceptance criteria, and outlier rejection are taken out of the hands of the user and are made a vital and integral part of the system. I justify this approach and demonstrate how a sensible mathematical approach must be tailored to the configuration of the instrument.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Negative values and variance functions: Implications for statistical analysis;Annals of Clinical Biochemistry: International Journal of Laboratory Medicine;2021-01-11
2. Multiple calibrator measurements improve accuracy and stability estimates of automated assays;Scandinavian Journal of Clinical and Laboratory Investigation;2016-01-19
3. Using the variance function to estimate limit of blank, limit of detection and their confidence intervals;Annals of Clinical Biochemistry: International Journal of Laboratory Medicine;2015-02-12
4. Calibration Curve Fitting;The Immunoassay Handbook;2013
5. Error models for immunoassays;Annals of Clinical Biochemistry: International Journal of Laboratory Medicine;2008-09