Affiliation:
1. Department of Pathology, University of South Alabama Medical Center, Mobile 36617
Abstract
Abstract
Five commercially available immunoassay techniques were compared for the detection of cannabinoid metabolites in urine at a positive cutoff of 100 micrograms/L. Numerical values, available for four of these techniques, were used to evaluate inter-assay equivalence at this threshold. The total number of positive results differed for each assay, with some individual samples exhibiting wide variations in numerical values. As a group, the radioimmunoassay systems gave a greater proportion of positive specimens than did enzyme-linked or fluorescence-polarization-based techniques. The single threshold of detection suggested by federal guidelines does not provide for equivalent screening of cannabinoids with these immunoassay techniques.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献