Abstract
Abstract
Background: Liquid chromatography-tandem mass spectrometry (LC-MS/MS) with electrospray ionization is robust and allows accurate measurement of both low- and high-molecular weight components of complex mixtures. We developed a LC-MS/MS method for the analysis of methylmalonic acid (MMA), a biochemical marker for inherited disorders of propionate metabolism and acquired vitamin B12 deficiency.
Methods: We added 1 nmol of the internal standard MMA-d3 to 500 μL of plasma or 100 μL of urine before solid-phase extraction. After elution with 18 mol/L formic acid, the eluate was evaporated, and butyl ester derivatives were prepared with 3 mol/L HCl in n-butanol at 65 °C for 15 min. For separation, we used a Supelcosil LC-18, 33 × 4.6 mm column with 60:40 (by volume) acetonitrile:aqueous formic acid (1 g/L) as mobile phase. The transitions m/z 231 to m/z 119 and m/z 234 to m/z 122 were used in the selected reaction monitoring mode for MMA and MMA-d3, respectively. The retention time of MMA was 2.2 min in a 3.0-min analysis, without interference of a physiologically more abundant isomer, succinic acid.
Results: Daily calibrations between 0.25 and 8.33 nmol in 0.5 mL exhibited consistent linearity and reproducibility. At a plasma concentration of 0.12 μmol/L, the signal-to-noise ratio for MMA was 40:1. The regression equation for our previous gas chromatography-mass spectrometry (GC-MS) method (y) and the LC-MS/MS method (x) was: y = 1.030x − 0.032 (Sy|x = 1.03 μmol/L; n = 106; r = 0.994). Inter- and intraassay CVs were 3.8–8.5% and 1.3–3.4%, respectively, at mean concentrations of 0.13, 0.25, 0.60, and 2.02 μmol/L. Mean recoveries of MMA added to plasma were 96.9% (0.25 μmol/L), 96.0% (0.60 μmol/L), and 94.8% (2.02 μmol/L). One MS/MS system used only overnight (7.5 h) replaced two GC-MS systems (30 instrument-hours/day) to run 100–150 samples per day, with reductions of total cost (supplies plus equipment), personnel, and instrument time of 59%, 14%, and 75%, respectively.
Conclusions: This method is well suited for large-scale MMA testing (≥100 samples per day) where a shorter analytical time is highly desirable. Reagents are less expensive than the anion-exchange/cyclohexanol-HCl method, and sample preparation of batches up to 100 specimens is completed in less than 8 h and is automated.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
89 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献