Affiliation:
1. Department of Physiology, University of Groningen, The Netherlands
Abstract
Abstract
Determination of the oxygen affinity of human whole blood with the help of blood-gas analyzers and multi-wavelength oximeters is compared with an accurate method for recording hemoglobin oxygen dissociation curves (Clin Chem 1982;28:1287-92). P50 (oxygen tension at half saturation; So2 = 50%) and Hill's n (delta log [So2/(1-So2)]/delta log Po2) were determined in blood of 24 healthy donors. Three slightly different procedures were used for determination of P50 on the basis of Po2, pH, Pco2, and So2 measured with each of four different blood-gas analyzer/oximeter combinations. These methods were not able to discriminate between high and low values for P50 within the normal reference interval, but never failed to detect the high oxygen affinity of blood stored for 12 days, reflected in low values of P50. The methods thus proved suitable for detecting clinically significant deviations in oxygen affinity such as occur in patients with, e.g., abnormal hemoglobins, anemias, or carbon monoxide poisoning. Determination of Hill's n by these methods did not produce useful results.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
15 articles.
订阅此论文施引文献