Affiliation:
1. Development Department, Syva Co., San Jose, CA 95161-9013
Abstract
Abstract
The major metabolite of cannabis found in urine, 11-nor-delta 9-tetrahydrocannabinol-9-carboxylic acid (delta 9-THC), is the compound most often used to calibrate cannabinoid immunoassays. The hydrophobic delta 9-THC molecule is known to adsorb to solid surfaces. This loss of analyte from calibrator solutions can lead to inaccuracy in the analytical system. Because the calibrators remain stable when not used, analyte loss is most probably caused by handling techniques. In an effort to develop an effective means of overcoming adsorption losses, we quantified cannabinoid loss from calibrators during the testing process. In studying handling of these solutions, we found noticeable, significant losses attributable to both the kind of pipette used for transfer and the contact surface-to-volume ratio of calibrator solution in the analyzer cup. Losses were quantified by immunoassay and by radioactive tracer. We suggest handling techniques that can minimize adsorption of delta 9-THC to surfaces. Using the appropriate pipette and maintaining a minimum surface-to-volume ratio in the analyzer cup effectively reduces analyte loss.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
17 articles.
订阅此论文施引文献