Influences of hemoglobin (Hb) variants (HbSS, HbCC, beta-thalassemia, HbAE, HbAS, HbAC, hereditary persistent HbF) and Hb derivatives (carbamylated- and acetylated-Hbs, Schiff base, and those formed in stored blood) on results of glyco-Hb assays by 102 laboratories using 16 different methods were investigated. Affinity chromatography shows deviating results only with homozygous Hb S and C. Correct interpretation of results from patients with decreased erythrocyte half-lives requires previous knowledge on this condition. Measurements of HbA1c by HPLC and electrophoresis are obviously unsuitable for homozygous hemoglobinopathies; for heterozygous hemoglobinopathies and Hb synthesis variants, HbA1c should be expressed as percentage of HbA0 + HbA1c; abnormal Hbs are usually recognized; both carbamylated- and acetylated-Hbs interfere and Schiff base must be eliminated. Except for stored blood, all Hb variants and derivatives gave erroneous results with disposable ion-exchange columns. Dako's immunoassay is not affected by Hb derivatives; glycated Hb variants are not recognized as glyco-Hb and percentages are consequently too low. Glyco-Hb by the immunoassay of Bayer (performed by one laboratory) is not affected by Hb variants and derivatives.