Affiliation:
1. Washington University School of Medicine, Department of Pathology, St. Louis, MO 63110
Abstract
Abstract
Myoglobin (Mb) is considered a useful marker for early detection of myocardial infarction and for monitoring cardiac reperfusion after thrombolytic therapy. We developed eight monoclonal antibodies to human cardiac Mb, characterized their epitopic reactivity, and determined which combinations of the antibodies are useful in two-site immunoassays. We configured two of the monoclonal antibodies in a one-step, two-site particle concentration fluorescence immunoassay (PCFIA) for measurement of Mb. The PCFIA has rapid kinetics of reaction, being complete in 15 min, and has a linear analytical range of 20-675 micrograms/L for human Mb. Although the PCFIA has a high-dose "hook" effect, this is of no analytical importance at concentrations of Mb less than or equal to 148,000 micrograms/L. The assay is not subject to interference from icterus (bilirubin less than or equal to 360 mg/L), has no cross-reaction with hemoglobin (less than or equal to 42 g/L), and may be performed with either plasma or serum in approximately 1 h. The intra- and interassay imprecisions (CV) of the method are less than 10% for concentrations of Mb within the normal range and less than 4% at higher concentrations. A comparison of the PCFIA with a commercial radioimmunoassay showed that results of the two assays correlate well (PCFIA = 0.88 x RIA + 18, r = 0.990, n = 171).
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
21 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献