Affiliation:
1. Department of Clinical Chemistry, Free University Hospital, Amsterdam, The Netherlands. klinlab@azvu.nl
Abstract
Abstract
The performance characteristics of two bone alkaline phosphatase (ALP; EC 3.1.3.1) assays, a wheat germ agglutinin (WGA) precipitation assay and a new immunoadsorption assay (IAA), were compared. The within- and between-run imprecision of the IAA (3.6-4.2% and 3.6-7.7%) was comparable with that of the WGA assay. The mean cross-reactivity with liver ALP appeared to be 4% in the WGA assay and 11% in the IAA. The reference ranges in a group of 155 healthy Caucasian (pre)pubertal schoolgirls were: 149-401 U/L (total ALP, 30 degrees C), 105-349 U/L (bone ALP, 30 degrees C, WGA assay), and 58-205 U/L (bone ALP, 25 degrees C, IAA). Comparison of the WGA assay (x) with the IAA (y) demonstrated a correlation coefficient of 0.95 [Deming regression equation: y = (0.56 +/- 0.01)x + (2.0 +/- 1.5); Sy[symbol: see text]x = 5.3 U/L]. Correlation studies of the WGA assay and the IAA results with total ALP demonstrated r = 0.98 and 0.96, respectively.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry