Abstract
Abstract
We describe enzymic fluorometric methods of automated analysis for glucose, lactate, pyruvate, 3-hydroxybutyrate, glycerol, and alanine in perchloric acid extracts of blood. Unmodified Technicon AutoAnalyzer II apparatus is used. The usual concentrations of all these metabolites can be measured in as little as 0.1 ml of blood from a fasting subject. Within-batch and between-batch coefficients of variation ranged from 0.4 to 4.4% for all metabolites except 3-hydroxybutyrate, for which CV's were higher for low concentrations. Analytical recovery of added metabolites ranged from 92 to 98%. Glucose, lactate, alanine, and 3-hydroxybutyrate are stable in perchloric acid extracts for at least 13 days at room temperature, and a year at -20 degrees C; pyruvate shows a 6--8% loss after 3 days and 52% by one year at -20 degrees C; glycerol concentrations were stable at -20 degrees C for at least 13 days. Blank fluorescence is found in perchloric acid extracts of blood, necessitating blank runs for pyruvate, 3-hydroxybutyrate, glycerol, and alanine. The systems are simple to use, relatively inexpensive to operate, and are recommended for any laboratory with high throughput of samples.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry