Affiliation:
1. Division of Laboratories of the Cedars Division of Cedars-Sinai Hospitals, Los Angeles, Cal. 90029
Abstract
Abstract
Natural gas was substituted for propane gas in the oxygen-fuel mixture supplying the Technicon AutoAnalyzer flame photometer. This was accomplished by the use of a 1/3-hp rotary pump and attached manifold, which compressed the gas from the city line and supplied it at a constant pressure to the burner module of the system. A method for the automatic analysis of serum Li was devised by using K as an internal standard and comparing the output of the Li-sensitive photo-cell with that of an attenuated K-sensitive cell. In addition, a simple procedure was incorporated into the routine determination of serum Na and K: samples of insufficient volume were diluted in a solution of known electrolyte concentration, which permitted their quantitation without changing the tubing manifold or other operating components of the system.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
7 articles.