Author:
Fogh-Andersen N,Christiansen T F,Komarmy L,Siggaard-Andersen O
Abstract
Abstract
We describe a new calcium ion-selective electrode for measurement of the substance concentration of free calcium ion [Ca2+] in the plasma phase of whole blood and in serum at 37 degrees C. A sample volume of 50 microliter suffices to obtain simultaneous values of pH and [Ca2+]. We found the within-series analytical standard deviation for serum to be 0.013 mmol/litre (CV, 1.1%) and day-to-day precision to be 0.022 mmol/litre (CV, 1.7%). The reference interval for [Ca2+] (at pH 7.40) in serum was found to be 1.184 +/- 0.054 mmol/litre (2 SD) from measurements on sera from 121 healthy blood donors. Measurements on capillary blood from 29 healthy volunteers gave a mean (+/- 2 SD) value for [Ca2+] (at pH 7.40) of 1.22 +/- 0.072 mmol/litre.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
67 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献