Affiliation:
1. Department of Pathology/SGHQLCC, Wilford Hall USAF Medical Center, Lackland AFB, San Antonio, TX 78236-5300
Abstract
Abstract
We evaluated the analytical and clinical performance of three "sensitive" immunoradiometric assay (IRMA) kits (Tandem-R TSH HS, Hybritech, Inc.; EchoClonal TSH, Bio-Rad; Coat-A-Count TSH IRMA, Diagnostic Products Corp.) for measurement of thyrotropin (TSH) and compared their performance against a "regular" IRMA (ARIA-HT, Becton-Dickinson) to determine whether these assays might eliminate the need to perform the thyroliberin (TRF) stimulation test. We concluded that the Tandem and EchoClonal kits may obviate the need to perform TRF stimulation tests in some patients. Using only the basal TSH concentration to predict the TSH response to TRF, we found all three sensitive TSH assays to be useful for detecting abnormal thyrotroph function. Dose-response, linearity, analytical recovery, and specificity were acceptable for all kits, but intra-assay precision at very low TSH concentrations and analytical sensitivity differed considerably among the kits. Using the EchoClonal assay, we established a normal reference interval for TSH of 0.4-4.6 milli-int. units/L.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献