Anodic stripping voltammetry procedure modified for improved accuracy of blood lead analysis.

Author:

Roda S M1,Greenland R D1,Bornschein R L1,Hammond P B1

Affiliation:

1. Department of Environmental Health, University of Cincinnati Medical Center, OH 45267-0056

Abstract

Abstract In evaluating the accuracy and reliability of blood lead (PbB) measurements with the Environmental Science Associates Model 3010A Trace Metal Analyzer, intralaboratory comparison demonstrated that use of the operating conditions recommended by the manufacturer resulted in consistently underestimated PbB concentrations less than 400 micrograms/L and overestimated PbB values greater than 400 micrograms/L. At PbB concentrations less than 50 micrograms/L, measured concentrations were often registered as negative results. However, these negative values could be replicated to within +/- 10 micrograms/L, indicating good precision of the method, but obviously not good accuracy. In addition, lower-than-expected lead (Pb) values were measured in samples containing increased concentrations of copper (Cu), such as may occur in pregnant women. We modified the procedure to eliminate these inaccuracies by substituting manual peak-height measurements for reliance on the integrator and digital display of the instrument. We established the accuracy of the modified procedure by using calibration standards previously quantified by isotope dilution-mass spectroscopy. A quality-control program for monitoring PbB analysis is also described.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3