Author:
McBride J H,Thibeault R V,Rodgerson D O
Abstract
Abstract
An immunoradiometric assay (Boots-Celltech's "Sucrosep"; IRMA) for thyrotropin (TSH) was evaluated and results were compared with those of our in-house RIA procedure. The IRMA had a sensitivity of 0.02 milli-int. unit/L. In addition it displayed excellent intra- and inter-batch precision, cross-reacted negligibly with other pituitary hormones, and appeared to be relatively free of matrix effects (although deionized water did give an apparent TSH concentration of 0.13 milli-int. unit/L). Over the range 1.2 to 64 milli-int. units/L, correlation between the IRMA (y) and RIA (x) was excellent: y = 1.16x - 1.69 (r = 0.98). The normal reference interval for the IRMA was 0.4 to 4.2 milli-int. units/L. For patients suspected of hyperthyroidism who had been subjected to the thyroliberin test, the IRMA more effectively differentiated between euthyroidism and hyperthyroidism. During testing with thyroliberin, all nonresponding hyperthyroid patients had initial baseline TSH concentrations of less than 0.02 milli-int. unit/L by IRMA. This sensitive assay represents an important advance in the ability to differentiate between euthyroidism and hyperthyroidism.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献