Affiliation:
1. All-Union Center for Phenylketonuria, Moscow, U.S.S.R
Abstract
Abstract
We adapted the method of McCaman and Robins for fluorometry of phenylalanine to a microplate assay for routine phenylketonuria screening. Sensitivity is 15 mumol/L for the plasma assay and 30 mumol/L for the dried blood-spot assay, with CV less than 10% for both assays. Results for human plasma by microplate assay correlated well (r = 0.99) with results of amino acid analyzer determination of phenylalanine. When measurements are performed in an automated reader, the microplate assay has considerable advantages over conventional measurements in cuvettes: smaller volumes of reagents and automation enabling high throughput and convenience. Because the described method is a quantitative one, we can postulate that, compared with the semiquantitative Guthrie inhibition bioassay, this microplate assay is more reliable, easier to perform, and about twofold less costly.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
42 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献