Abstract
Abstract
We describe a reliable gas-chromatographic analysis for unchanged phencyclidine in human plasma, with use of a nitrogen detector. The assay is sufficiently sensitive to permit measurement of as little as 5 microgram of the drug per liter, in 2 mL of plasma. Phencyclidine is extracted from plasma at pH 10.5 into hexane/iso-amyl alcohol containing meperidine as the internal standard, back-extracted into dilute HCl, and re-extracted into hexane/iso-amyl alcohol after alkalinization of the HCl. The coefficients of variation for the analysis at respective concentrations of 100 and 200 microgram/L are: within-run 4.0% and 3.4%; between-run, 2.8% and 5.2%. This procedure, used in our laboratory for more than one year, has been applied to plasma from numerous individuals with nonfatal phencyclidine intoxication. Some representative examples are given.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
22 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献