Author:
Annesley T,Matz K,Balogh L,Clayton L,Giacherio D
Abstract
Abstract
This liquid-chromatographic assay requires 0.2 to 0.5 mL of whole blood, avoids the use of diethyl ether, and consumes only 10 to 20% of the solvents used in prior methods. Sample preparation involves an acidic extraction with methyl-t-butyl ether, performed in a 13 X 100 mm disposable glass tube, then a short second extraction of the organic phase with sodium hydroxide. After evaporation of the methyl-t-butyl ether, chromatography is performed on an "Astec" 2.0-mm (i.d.) octyl column. We compared results by this procedure with those by use of earlier larger-scale extractions and their respective 4.6-mm (i.d.) columns; analytical recoveries of cyclosporins A and D were comparable with previous findings and results for patients' specimens were equivalent, but the microbore columns provided greatly increased resolution and sensitivity.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献