Affiliation:
1. Centralized Department of Clinical Chemistry and Hematology, St. Elisabeth Hospital, Tilburg, The Netherlands
Abstract
Abstract
Severe hyperglycemia can result in falsely high results for mean cell (erythrocyte) volume (MCV), which will also cause false results for erythrocyte indices calculated on the basis of MCV. Falsely high MCV results were obtained with the Technicon H1 and H2 analyzers and (to a lesser extent) with the Coulter T660. The H analyzers were more susceptible to this interference than was the Coulter T660. This difference in sensitivity of MCV to hyperglycemia can be explained by the use of sodium dodecyl sulfate in the Technicon erythrocyte diluent and by differences in incubation times. In severe hyperglycemia, results for MCV, mean cell hemoglobin concentration, and hematocrit obtained from electronic cell counters, especially Technicon H systems, are unreliable.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Megaloblastic Anemia;Decision Making Through Problem Based Learning in Hematology;2024
2. Autoimmune Hemolytic Anemia;Decision Making Through Problem Based Learning in Hematology;2024
3. Hématologie;Revue Francophone des Laboratoires;2016-02
4. Sensitivity of Hematocrit to Osmotic Effects Induced by Changes in Dialysate Conductivity;ASAIO Journal;2015-09
5. Hypo- and hypernatremia results in inaccurate erythrocyte mean corpuscular volume measurementin vitro, when using Sysmex XE 2100;Scandinavian Journal of Clinical and Laboratory Investigation;2015-08-12