Affiliation:
1. Department of Pathology, White Memorial Medical Center, Los Angeles, CA 90033
Abstract
Abstract
We have designed a re-usable dialysis cell and a complex dialysis buffer, with which undiluted serum samples can be dialyzed with minimal changes in their serum matrix. Dialysate thyroxin (free T4) is then measured by a sensitive RIA for T4. The range of reportability was 2-128 ng/L, the normal range was 8-27 ng/L, and the interassay CV was 7%. Free T4 concentrations in various disorders were as follows: hyperthyroidism, 32-478 ng/L; in both excess thyroxin-binding globulin (TBG) and familial dysalbuminemic hyperthyroxinemia, 9-27 ng/L; primary hypothyroidism, less than 2-7 ng/L; central hypothyroidism, 4-6 ng/L; severe TBG deficiency, 9-25 ng/L; hypothyroxinemias of nonthyroidal illness, 8-35 ng/L. With this free-T4 assay, which is adaptable to clinical laboratory use, one can differentiate hyperthyroidism from the major euthyroid hyperthyroxinemias and hypothyroidism from the major euthyroid hypothyroxinemias.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
106 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献