Decreased signal in Emit assays of drugs of abuse in urine after ingestion of aspirin: potential for false-negative results

Author:

Wagener R E1,Linder M W1,Valdes R1

Affiliation:

1. Department of Pathology, University of Louisville School of Medicine, University of Louisville Hospital, KY 40292

Abstract

Abstract During routine drug analysis with the Syva d.a.u. Emit immunoassays we observed a high frequency of urines with lower rates of changes in absorbance (delta A R) than the rate for a drug-free urine calibrator. Many of these urines contained salicylates. Among 40 urines with apparent salicylate concentrations between 15 and 420 mg/dL tested for benzoylecgonine (BE), 20 had delta A R < -4 (range +2 to -28 mA/min). The rates decreased with increasing salicylate: delta A R = -0.057 x (salicylate, mg/dL) -0.22 mA/min (r = 0.85, n = 40, P < 0.01). Urines from 100 control subjects (no salicylate) had mean +/- SD delta A R values of -1.05 +/- 2.2 mA/min (range +3 to -7; only two were < -4 mA/min). Although direct addition of salicylic acid (200 mg/dL) to urine specimens did not reproduce the negative bias, ingestion of aspirin (acetylsalicylic acid) did by -0.09 mA/min per 1 mg/dL (72.4 mumol/L) salicylate. Negative biases observed for other Emit d.a.u. assays after salicylate ingestion lead us to conclude that ingestion of therapeutic doses of aspirin may cause false-negative results for drug screens in urines by this technology.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

Cited by 17 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3