Affiliation:
1. Oak Ridge National Laboratory, Oak Ridge, Tenn. 37830. (Operated for the U. S. AEC by Union Carbide Corp.)
Abstract
Abstract
A series of miniature photometers has been developed for use as on-stream detectors in systems for liquid chromatography. All of the detectors feature dual-beam, dual-wavelength operation without mechanical or electrical switching of either wavelength or signal. The two types of detectors originally developed—ultraviolet (254 and 280 nm) and colorimetric (any two choices of visible wavelengths)—have been integrated into a single modular instrument that can be converted from ultraviolet to use with visible wavelengths by substituting source and filter modules. Both straight-through vertical quartz tube flow cells of 3-mm i.d. and Z-cell modules with 10-mm light pathlengths are available. A single-unit power-source and rangeswitching box is used with any configuration of the detector.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献