Routine Radioimmunoassay of Plasma Testosterone, and Results for Various Endocrine Disorders

Author:

Wong Pui-Yuen1,Wood Donald E1,Johnson Thomas1

Affiliation:

1. Toronto General Hospital, 101 College St., Toronto, Ontario, Canada M5G 1L7

Abstract

Abstract We describe a modification of published methods for radioimmunoassay of plasma testosterone. This simpler method involves no chromatographic steps, and the necessary reagents, including tritiated testosterone and testosterone anti-serum, are commercially available. A digital computer is used for the calculation. Without difficulty, a technician can complete 100 assays in three working days. Mean testosterone concentrations (± 2 SD) in the plasmas of 21 normal men and 26 women with a normal menstrual cycle were 684 ± 300 and 45 ± 2O ng/dl, respectively. Within- and between-assay precision (coefficient of variation) were 5.2% (n = 29) and 6.7% (n = 26), respectively. We have assayed more than 1000 samples during the past year. We give data on the concentrations of testosterone in plasma of patients with various endocrine disorders.