Affiliation:
1. Department of Pathology, University of Maryland, School of Medicine, Baltimore 21201
Abstract
Abstract
This two-site IRMA includes specific monoclonal antibodies for measuring skeletal alkaline phosphatase (B-ALP) in human serum. Assay calibration is based on mass units (micrograms per liter) and was established with purified B-ALP from a human osteosarcoma cell line, SAOS-2. Precision studies demonstrated intra- and interassay CVs of 3-5% and 5-7%, respectively. Relative reactivity studies showed that the assay has a sevenfold preference for detecting B-ALP compared with the liver isoenzyme in serum. The normal reference interval for 478 healthy adults was 5-22 micrograms/L. Method comparison studies showed good correlation between this B-ALP assay (y) and commercially available electrophoretic methods (x) (y = 0.3540x + 20.5, R2 = 0.929) in a pagetic population. Temporal profiles for total ALP, this IRMA B-ALP assay, and B-ALP by electrophoresis in three pagetic patients were parallel. We conclude that this assay demonstrates good analytical performance and would be useful for the clinical assessment of metabolic bone disorders.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
58 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献