Affiliation:
1. Department of Health Sciences, Hong Kong Polytechnic, Kowloon
Abstract
Abstract
We describe a simple enzyme-linked immunosorbent assay (ELISA) in which microtiter plates are used for determining the free thyroxin concentration (FT4) in serum. Only commercially available chemicals and reagents are needed, including the antiserum. The working range is 1 to 60 ng/L. Turnaround time is about 4 h. Within-run coefficients of variation (CV) for FT4 concentrations of 9.0, 19.0, and 35.0 ng/L were less than 5%; between-run CVs were 10% to 11%. Results by a radioimmunoassay method for 150 sera correlated well (r = 0.922).
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献