Author:
Geiseler D,Chodha P,Ekins R
Abstract
Abstract
In binding assays for determination of free (non-protein-bound) hormones in serum or plasma, the influence of the measuring system on the original analyte concentration in the sample must be considered. In one- or single-step free-hormone immunoassays, the labeled analyte or analog-tracer not only is bound to the antibody, it also is bound, to some extent, to serum proteins. The dependence of the assay response on two unknown variables--the concentration of free analyte and the binding potential of serum for the tracer--introduces a bias between the actual (original) and measured hormone concentrations. The significance of this protein effect is described by mathematical modeling of the analyte-protein distribution in the assay system. The theoretical consideration is validated by a clearly defined one-step assay system for measurement of free-thyroxin concentration, with a labeled thyroxin-immunoglobulin conjugate used as tracer.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
7 articles.