Zucchelli G C,Pilo A,Chiesa M R,Masini S,Baccini C
Abstract
Abstract
Between-laboratory agreement of CEA determinations collected in a multicenter study has been substantially improved (CV decreased from 60.4 to 21.6%) by converting the results from mass concentration units (ng/mL) of local kit standards to int. units/L of the international standard first IRP 73/601. Conversion factors for the kits most used were computed, during the same survey, from results of two analytical-recovery experiments in which control samples, supplemented with the international standard, were analyzed by participating laboratories (n = 96).
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry