Affiliation:
1. Chemistry Department, Research Section, Kiess Instruments, Inc., 8768 S.W. 131st St., Miami, Fla. 33156
Abstract
Abstract
The glucose oxidase—peroxidase procedure for determining glucose in serum or plasma has been modified by changing the assay pH to 5.5, altering the ratio of glucose oxidase to peroxidase, and using a soluble chromogen, p-diphenylamine sulfonate, to prepare a single complete reagent that develops stable color within 2 min without deproteinization. Most interferences were inconsequential. Serum that clearly is icteric or hemolytic is Somogyi-precipitated. Recovery was 99% and the CV was 3.3%. Correlation with results obtained by the neocuproine procedure, with use of the SMA 12/60, averaged 12 mg/dl lower values (r2 = 0.98).
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
33 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献