Potential False-Negative Nucleic Acid Testing Results for Severe Acute Respiratory Syndrome Coronavirus 2 from Thermal Inactivation of Samples with Low Viral Loads

Author:

Pan Yang123,Long Luyao456,Zhang Daitao123,Yuan Tingting7,Cui Shujuan123,Yang Peng123,Wang Quanyi123,Ren Simei456

Affiliation:

1. Institute for Infectious Disease and Endemic Disease Control, Beijing Center for Disease Prevention and Control, Beijing, P. R. China

2. Research Centre for Preventive Medicine of Beijing, Beijing, China

3. School of Public Health, Capital Medical University, Beijing, China

4. National Center for Clinical Laboratories, Beijing Hospital, National Center of Gerontology, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing, China

5. Graduate School, Chinese Academy of Medical Sciences, Peking Union Medical College, Beijing, China

6. Beijing Engineering Research Center of Laboratory Medicine, Beijing Hospital, Beijing, China

7. Department of Clinical Laboratory, Beijing Hospital, National Center of Gerontology, Beijing, China

Abstract

Abstract Background Coronavirus disease-2019 (COVID-19) has spread widely throughout the world since the end of 2019. Nucleic acid testing (NAT) has played an important role in patient diagnosis and management of COVID-19. In some circumstances, thermal inactivation at 56°C has been recommended to inactivate severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) before NAT. However, this procedure could theoretically disrupt nucleic acid integrity of this single-stranded RNA virus and cause false negatives in real-time polymerase chain reaction (RT-PCR) tests. Methods We investigated whether thermal inactivation could affect the results of viral NAT. We examined the effects of thermal inactivation on the quantitative RT-PCR results of SARS-CoV-2, particularly with regard to the rates of false-negative results for specimens carrying low viral loads. We additionally investigated the effects of different specimen types, sample preservation times, and a chemical inactivation approach on NAT. Results Our study showed increased Ct values in specimens from diagnosed COVID-19 patients in RT-PCR tests after thermal incubation. Moreover, about half of the weak-positive samples (7 of 15 samples, 46.7%) were RT-PCR negative after heat inactivation in at least one parallel testing. The use of guanidinium-based lysis for preservation of these specimens had a smaller impact on RT-PCR results with fewer false negatives (2 of 15 samples, 13.3%) and significantly less increase in Ct values than heat inactivation. Conclusion Thermal inactivation adversely affected the efficiency of RT-PCR for SARS-CoV-2 detection. Given the limited applicability associated with chemical inactivators, other approaches to ensure the overall protection of laboratory personnel need consideration.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

Reference24 articles.

1. Early epidemiological analysis of the coronavirus disease 2019 outbreak based on crowdsourced data: a population-level observational study;Sun;Lancet Digit Health,2020

2. Crowdsourcing data to mitigate epidemics;Leung;Lancet Digit Health

3. Clinical course and outcomes of critically ill patients with SARS-CoV-2 pneumonia in Wuhan, China: a single-centered, retrospective, observational study;Yang;Lancet Respir Med,2020

4. A pneumonia outbreak associated with a new coronavirus of probable bat origin;Zhou;Nature,2020