Author:
Sheehan M,Hertel R H,Kelly C T
Abstract
Abstract
We describe a combined gas-chromatographic mass-spectrometric method for determination of theophylline in whole blood. With use of a probability-based matching approach, the extensive gas-chromatographic interference did not affect mass-spectrometric quantitation of theophylline. Concentrations expected and found were linearly related from 1 to 60 mg/liter; the run-to-run coefficient of variation was less than 8%, within-run less than 6%. Some other xan-thines and drugs frequently co-administered with theophylline did not affect results. Only 0.2 ml of blood is required, which makes the procedure useful for monitoring theophylline in newborns and pediatric populations.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献