Revised calibration of the Reflotron cholesterol assay evaluated.

Author:

Boerma G J1,van Gorp I1,Liem T L1,Leijnse B1,Bellm J1,Carstensen C A1

Affiliation:

1. Department of Clinical Chemistry, Academic Hospital Rotterdam, The Netherlands

Abstract

Abstract We evaluated the Boehringer Mannheim (B.M.) Reflotron Total Cholesterol "dry-chemistry" method after its recalibration in 1987. Reports in the literature up to 1986-1987 of a negative bias (up to -10%) in the method prompted a revision of the factory-set calibration of the Reflotron. For this, B.M. prepared a new set of calibrators with 12 different concentrations of cholesterol. We checked in two ways whether accuracy had been achieved: (a) The values assigned to the calibrators by B.M. were checked with the manual Abell-Kendall Reference Method (MAK) performed in an official Reference Center. These were shown to be correct. (b) Concurrently, a direct comparison was made by analyzing 200 fresh samples of human serum. Reflotron cholesterol values obtained for these samples proved to be accurate, meeting the current World Health Organization/Centers for Disease Control criterion of maximum bias less than or equal to 5%. Orthogonal regression analysis yielded the following correlation: Reflotron = 0.985 MAK + 0.238 mmol/L (y = ax + b). Reflotron mean = 6.26 mmol/L; MAK mean = 6.09 mmol/L. SDa = 0.015 mmol/L; SDb = 0.120 mmol/L, and r = 0.989.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

Cited by 13 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Determination of cholesterol in blood. Part 2;Review Journal of Chemistry;2016-04

2. Effects of a school-based intervention to reduce cardiovascular disease risk factors in elementary-school children: The Cardiovascular Health in Children (CHIC) Study;The Journal of Pediatrics;1996-06

3. Reflotron Cholesterol Measurement in General Practice: Accuracy and Detection of Errors;Annals of Clinical Biochemistry: International Journal of Laboratory Medicine;1994-11

4. Reflotron;Dry Chemistry - Analysis with Carrier-Bound Reagents;1993

5. Performance of the Reflotron and the Seralyzer in primary health care;Scandinavian Journal of Clinical and Laboratory Investigation;1993-01

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3