Abstract
Abstract
An automated method has been developed in which a new o-toluidine reagent is used for the microanalysis of glucose. The reagent is 1.3 times as sensitive as the o-toluidine reagent commonly used in manual procedures. The reagent does not significantly alter the life expectancy of manifold tubing made of polyvinyl chloride. The automation includes a dialysis step and possesses the desirable features of the manual methods, including the greater specificity as compared with methods based on reducing properties. Results of this automated method obey Beer’s law, are very reproducible (the coefficient of variation is approximately 1.5%), and are not significantly different from those obtained manually. The method requires 75 µl of sample.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry